• Media type: E-Article
  • Title: Adjoint-based shape optimization for the minimization of flow-induced hemolysis in biomedical applications
  • Contributor: Bletsos, Georgios [Author]; Kühl, Niklas [Author]; Rung, Thomas [Author]
  • Corporation: Technische Universität Hamburg ; Technische Universität Hamburg, Institut für Fluiddynamik und Schiffstheorie
  • Published: 2021
  • Published in: Engineering applications of computational fluid mechanics ; 15(2021), 1, Seite 1095-1112
  • Language: English
  • DOI: 10.15480/882.3659; 10.1080/19942060.2021.1943532
  • Identifier:
  • Keywords: Computational fluid dynamics(CFD) ; adjoint-based shape optimization ; biomedical design ; hemolysis minimization
  • Origination:
  • Footnote: Sonstige Körperschaft: Technische Universität Hamburg
    Sonstige Körperschaft: Technische Universität Hamburg, Institut für Fluiddynamik und Schiffstheorie
  • Description: This paper reports on the derivation and implementation of a shape optimization procedure for the minimization of hemolysis induction in biomedical devices. Hemolysis is a blood damaging phenomenon that may occur in mechanical blood-processing applications where large velocity gradients are found. An increased level of damaged blood can lead to deterioration of the immune system and quality of life. It is, thus, important to minimize flow-induced hemolysis by improving the design of next-generation biomedical machinery. Emphasis is given to the formulation of a continuous adjoint complement to a power-law hemolysis prediction model dedicated to efficiently identifying the shape sensitivity to hemolysis. The computational approach is verified against the analytical solutions of a benchmark problem and computed sensitivity derivatives are validated by a finite differences study on a generic 2D stenosed geometry. The application included addresses a 3D ducted geometry which features typical characteristics of biomedical devices. An optimized shape, leading to a potential improvement in hemolysis induction up to 22%, is identified. It is shown, that the improvement persists for different, literature-reported hemolysis-evaluation parameters.
  • Access State: Open Access
  • Rights information: Attribution (CC BY)