• Media type: E-Article
  • Title: A Case of Implant Failure in Partial Wrist Fusion Applying Magnesium-Based Headless Bone Screws
  • Contributor: Wichelhaus, Alice; Emmerich, Judith; Mittlmeier, Thomas
  • imprint: Hindawi Limited, 2016
  • Published in: Case Reports in Orthopedics
  • Language: English
  • DOI: 10.1155/2016/7049130
  • ISSN: 2090-6749; 2090-6757
  • Keywords: Insect Science ; Ecology ; Ecology, Evolution, Behavior and Systematics
  • Origination:
  • Footnote:
  • Description: <jats:p>This article presents a case of implant failure resulting in mechanical instability of a scaphotrapezotrapezoideal arthrodesis using magnesium-based headless bone screws. During revision surgery osteolysis surrounding the screws was observed as well as degraded screw threads already in existence at 6 weeks after implantation. The supposed osseous integration attributed to magnesium-based screws could not be reproduced in this particular case. Thus, it can be reasoned that the use of magnesium-based screws for partial wrist arthrodesis cannot be encouraged, at least not in dual use.</jats:p>
  • Access State: Open Access