• Media type: E-Article
  • Title: Spurious Detection of Ventricular Fibrillation By a Transvenous Implantable Cardioverter Defibrillator During Percutaneous Transluminal Coronary Angioplasty
  • Contributor: RIZK, ANTOINE B.; CHANDLER, JOHN P.; SCHOENFELD, MARK H.; MARIEB, MARK A.
  • Published: Wiley, 1997
  • Published in: Pacing and Clinical Electrophysiology, 20 (1997) 1, Seite 136-137
  • Language: English
  • DOI: 10.1111/j.1540-8159.1997.tb04826.x
  • ISSN: 0147-8389; 1540-8159
  • Keywords: Cardiology and Cardiovascular Medicine ; General Medicine
  • Origination:
  • Footnote:
  • Description: <jats:p>This case describes another example of inappropriate implantable defibrillator shock. We also show that this phenomenon resulted from electromagnetic current traveling down a coronary wire during coronary angioplasty.</jats:p>