Titel:
Dose rate effects in radiochromic leuco crystal violet dosimeters
Beteiligte:
Jordan, Kevin
Erschienen:
IOP Publishing, 2022
Erschienen in:
Journal of Physics: Conference Series, 2167 (2022) 1, Seite 012008
Sprache:
Nicht zu entscheiden
DOI:
10.1088/1742-6596/2167/1/012008
ISSN:
1742-6588;
1742-6596
Entstehung:
Anmerkungen:
Beschreibung:
Abstract Radiochromic dosimeters based on leuco crystal violet (LCV) in aqueous solutions and gelatin hydrogels were evaluated for dose and dose rate performance. Optical transmission measurements were performed on samples in custom 10 cm long cuvettes in order to have measurable signals below 10 Gy. Standard gel formulations with 0.7 mM concentration of anionic surfactant sodium dodecyl sulfate (SDS) had a dose rate effect of increasing sensitivity with increasing dose rate of nearly 15% from 1 to 10 Gy/min at 20 Gy. Solutions of LCV and trichloracetic acid also showed similar dose rate effects. Adding LCV increased dose rate effect. Adding SDS, lowered sensitivity but maintained dose rate effect. Adding 0.4% gelatin also lowered sensitivity but reversed dose rate effect. For gels, lowering the pH reduced both sensitivity and dose rate effect. These trends are consistent with LCV solubility contributing to dose rate effects. These data indicate there are several factors related to dose rate effects in leuco crystal violet dosimeters and formulations designed to maximize sensitivity may not minimize dose rate effects.