• Medientyp: E-Artikel
  • Titel: High-resolution phonocardiogram parameters
  • Beteiligte: Zemlyakov, I Yu; Zdanov, D S; Kosteley, Y V; Golobokova, E V; Dikman, E Yu; Bureev, A Sh
  • Erschienen: IOP Publishing, 2021
  • Erschienen in: IOP Conference Series: Materials Science and Engineering, 1019 (2021) 1, Seite 012025
  • Sprache: Nicht zu entscheiden
  • DOI: 10.1088/1757-899x/1019/1/012025
  • ISSN: 1757-8981; 1757-899X
  • Entstehung:
  • Anmerkungen:
  • Beschreibung: <jats:title>Abstract</jats:title> <jats:p>The article describes the results of studying and analyzing phonocardiograms (PCGs) obtained during a physiological experiment with Blu-ray standard equipment. It provides the findings of a spectral and spectral-time analysis for signals with a sampling frequency of 10, 44.1 and 192 kHz. It shows that the differences in the PCG spectra of identical signals are unreliable. The article specifies the onset and disappearance moments of the harmonic components of heart sounds. It also provides recommendations on the sampling frequency and bit resolution of digitized PCG signals for telemetric systems.</jats:p>
  • Zugangsstatus: Freier Zugang