• Medientyp: E-Artikel
  • Titel: Highly Stable Glassy Carbon Interfaces for Long-Term Neural Stimulation and Low-Noise Recording of Brain Activity
  • Beteiligte: Vomero, Maria; Castagnola, Elisa; Ciarpella, Francesca; Maggiolini, Emma; Goshi, Noah; Zucchini, Elena; Carli, Stefano; Fadiga, Luciano; Kassegne, Sam; Ricci, Davide
  • Erschienen: Springer Science and Business Media LLC, 2017
  • Erschienen in: Scientific Reports
  • Sprache: Englisch
  • DOI: 10.1038/srep40332
  • ISSN: 2045-2322
  • Entstehung:
  • Anmerkungen:
  • Beschreibung: <jats:title>Abstract</jats:title><jats:p>We report on the superior electrochemical properties, <jats:italic>in-vivo</jats:italic> performance and long term stability under electrical stimulation of a new electrode material fabricated from lithographically patterned glassy carbon. For a direct comparison with conventional metal electrodes, similar ultra-flexible, micro-electrocorticography (μ-ECoG) arrays with platinum (Pt) or glassy carbon (GC) electrodes were manufactured. The GC microelectrodes have more than 70% wider electrochemical window and 70% higher CTC (charge transfer capacity) than Pt microelectrodes of similar geometry. Moreover, we demonstrate that the GC microelectrodes can withstand at least 5 million pulses at 0.45 mC/cm<jats:sup>2</jats:sup> charge density with less than 7.5% impedance change, while the Pt microelectrodes delaminated after 1 million pulses. Additionally, poly(3,4-ethylenedioxythiophene)-poly(styrenesulfonate) (PEDOT-PSS) was selectively electrodeposited on both sets of devices to specifically reduce their impedances for smaller diameters (&lt;60 μm). We observed that PEDOT-PSS adhered significantly better to GC than Pt, and allowed drastic reduction of electrode size while maintaining same amount of delivered current. The electrode arrays biocompatibility was demonstrated through <jats:italic>in-vitro</jats:italic> cell viability experiments, while acute <jats:italic>in vivo</jats:italic> characterization was performed in rats and showed that GC microelectrode arrays recorded somatosensory evoked potentials (SEP) with an almost twice SNR (signal-to-noise ratio) when compared to the Pt ones.</jats:p>
  • Zugangsstatus: Freier Zugang