• Medientyp: E-Artikel
  • Titel: Acoustic measures of low-frequency noise in extended high-frequency audiometry
  • Beteiligte: Schmuziger, Nicolas; Brechbuehl, Matthias; Probst, Rudolf
  • Erschienen: Acoustical Society of America (ASA), 2007
  • Erschienen in: The Journal of the Acoustical Society of America
  • Sprache: Englisch
  • DOI: 10.1121/1.2437848
  • ISSN: 0001-4966; 1520-8524
  • Schlagwörter: Acoustics and Ultrasonics ; Arts and Humanities (miscellaneous)
  • Entstehung:
  • Anmerkungen:
  • Beschreibung: <jats:p>A very high signal-to-noise ratio is required for equipment designed for extended high-frequency audiometry because listeners with almost no hearing ability in the extended high-frequency range may have normal hearing sensitivity in the lower frequencies. Two commercially available systems designed for pure-tone audiometry were evaluated both in the conventional and extended high-frequency range. Unwanted lower frequency signals greater than the noise floor occurred predominantly at presentation levels of approximately 110dB SPL or higher. Test tones in the extended high-frequency range should be restricted to levels that are not associated with lower frequency noise.</jats:p>