• Medientyp: E-Artikel
  • Titel: Performance of the Abbott RealTime HCV Genotype II RUO Assay
  • Beteiligte: Shinol, Rebecca C.; Gale, Howard B.; Kan, Virginia L.
  • Erschienen: American Society for Microbiology, 2012
  • Erschienen in: Journal of Clinical Microbiology
  • Sprache: Englisch
  • DOI: 10.1128/jcm.01249-12
  • ISSN: 0095-1137; 1098-660X
  • Schlagwörter: Microbiology (medical)
  • Entstehung:
  • Anmerkungen:
  • Beschreibung: <jats:title>ABSTRACT</jats:title> <jats:p> The Abbott RealTime HCV Genotype II RUO (research use only) assay was evaluated using the automated Abbott RealTime <jats:italic>m</jats:italic> 2000 system. Concordance was 98% (81/83 samples) with samples previously typed by the Versant HCV Genotype 2.0 RUO system with manual extraction. The total assay time was reduced from 10.5 to 6.0 h and hands-on time from 13 to 4 min/patient sample. </jats:p>
  • Zugangsstatus: Freier Zugang